MUE19 Multi Channels Ultrasonic Pressure Monitor

Overview

Front panel

 

Rear panel

Front panel

* Monitor Meter CH A~D 

* Gain Range Selector

* Alarm Set Indicator CH A~D

* Alarm Set Knob CH A~D

* Power Switch

 

 

Rear panel

* test Rods Socket  CH A~D

* Data Output Socket

* Data Output Terminals

* Power Supply Socket

 

instructions

[1]

First installed the L type test rods in the fixed positions of the vibration tank.

Put the end of test rods into the water around 3cm. The depth could adjust as needed.

The depth should be recorded for reference. The data should tested and recorded at the same depth to reduce the error.

[2]

Plug the A~D test rod cables into the input sockets of the corresponding channel of the unit.

[3]

Select the appropriate level range as required, such as:

Low frequency, 15KHz~60KHz, could set at 0~50mV range;

Mid frequency, 60KHz~250KHz, could set at 0~100mV range;

Hi frequency, 250KHz~3MHz, could set at 0~500mV range.

[4]

Connect the power cord to the AC outlet. Check the supply voltage is correct. Turn on the power.

Check the selected level range of each channel is correct. If OK, can start the data reading and recording.

[5]

The scale 0~10 of the alarm set knob are indicated the value 0~100% of the selected level range.

For example, level range is 0~50mV, Alarm set at 25mV, then, the alarm set knob should set at 5.0 position.

Or level range is 0~500mV, alarm set at 300mV, then, the alarm set knob should set at 6.0 position.  And so on.

[6]

If the detected signal level of test rod is lower than the alarm setting, then the alarm be triggered, the red alarm indicator is on and the alarm output socket output DC12V.

If the alarm set knob set at zero position, this setting is same as switch off the alarm. The red alarm indicator should off.

[7]

After use, please turn off the power.

 