MUE19 Multi Channels Ultrasonic Pressure Monitor
Overview |
Front panel
Rear panel |
Front panel
* Monitor Meter CH A~D
*
Gain
Range
Selector
* Alarm Set
Indicator CH A~D
* Alarm Set Knob CH A~D
* Power Switch
Rear
panel
* test Rods Socket
CH A~D
* Data Output Socket
* Data Output
Terminals
* Power Supply Socket
|
instructions |
[1] |
First installed the L
type test rods in the
fixed positions of the
vibration tank.
Put the end of test rods
into the water around
3cm. The depth could
adjust as needed.
The depth should be
recorded for reference.
The data should tested and
recorded
at the same depth to
reduce the error.
|
[2] |
Plug the A~D test rod
cables into the input
sockets of the
corresponding
channel
of the unit. |
[3] |
Select the appropriate
level range as required,
such as:
Low frequency,
15KHz~60KHz, could set
at 0~50mV range;
Mid frequency,
60KHz~250KHz, could set
at 0~100mV range;
Hi frequency,
250KHz~3MHz, could set
at 0~500mV range. |
[4] |
Connect the power
cord to the AC outlet.
Check the supply
voltage is correct.
Turn on the power.
Check the selected level
range of each channel is
correct. If OK, can
start the data reading
and recording. |
[5] |
The scale 0~10 of the
alarm
set knob are
indicated the value 0~100% of
the
selected level range.
For example,
level range is 0~50mV,
Alarm set at
25mV, then, the
alarm set knob should
set at 5.0 position.
Or level range is
0~500mV, alarm set at
300mV, then, the
alarm
set knob should
set at 6.0 position.
And so on.
|
[6] |
If the detected signal
level of test rod is lower than the
alarm setting, then the
alarm be
triggered, the red alarm
indicator is on
and the alarm output
socket output
DC12V.
If the
alarm
set knob set at
zero
position,
this setting is same as
switch off the alarm.
The red alarm indicator
should off.
|
[7] |
After use,
please turn off the
power.
|
Last Update:2022-06-18 |